Field dissemination of the value of the quantity method of the spectroradiometer in visible-to-shortwave infrared

  • role: First author第一作者
  • Affiliation:

    Division of Metrology in Optics and Laser, National Institute of Metrology, Beijing 100029, China

  • Email:chengqt@nim.ac.cn
  • Introduction:E-mail chengqt@nim.ac.cn
CHENG Qiutong1,  
  • Affiliation:

    Division of Metrology in Optics and Laser, National Institute of Metrology, Beijing 100029, China

LI Ling1,  
  • Affiliation:

    Division of Metrology in Optics and Laser, National Institute of Metrology, Beijing 100029, China

DAI Caihong1,  
  • role: Corresponding author通信作者
  • Affiliation:

    Division of Metrology in Optics and Laser, National Institute of Metrology, Beijing 100029, China

  • Email:wuzf@nim.ac.cn
  • Introduction:E-mail wuzf@nim.ac.cn
WU Zhifeng1*,  
  • Affiliation:

    Key Laboratory of Quantitative Remote Sensing Information Technology, Aerospace Information Research Institute, Chinese Academy of Sciences, Beijing 100094, China

GAO Caixia2,  
  • Affiliation:

    Division of Metrology in Optics and Laser, National Institute of Metrology, Beijing 100029, China

HE Sijie1,  
  • Affiliation:

    Division of Metrology in Optics and Laser, National Institute of Metrology, Beijing 100029, China

WANG Yanfei1

résumé

Field spectroradiometers need to be traced back to a unified benchmark, namely, the international system (SI) of units, to ensure the accuracy and consistency of different field spectroradiometers. Considering that reports focusing on the accuracy of field radiometric calibration of spectroradiometers are lacking, this study investigates the field dissemination of the value of the quantity method.Uncertainty evaluation methods, such as the model of the relationship between wavelength and spectral radiance, the stray light correction model, and the temperature correction model, are provided for the transfer spectroradiometer. Then, the radiometric quantity is transferred from the laboratory measurement standard to the transfer spectroradiometer and then to the field instruments. By analyzing the uncertainty components in the whole process, a laboratory-field dissemination method of the value of the quantity chain is established, and the field spectroradiometer is traced to the SI unit uninterruptedly. In accordance with the uncertainty propagation law, the uncertainties of the transfer spectroradiometer and field spectroradiometer are obtained. Experimental results show that the uncertainty of the transfer spectroradiometer is 1.7%—2.3% (k = 1) from 380 nm to 2400 nm, and the uncertainty of the field spectroradiometer is 1.9%—2.5% (k = 1) from 380 nm to 2400 nm.The paper systematically introduces the method of field dissemination of the value of the quantity method. During the field calibration process, the uncertainty and characteristics of the standard transfer spectroradiometer are investigated. By analyzing the field calibration uncertainty in the whole process, the field spectroradiometer can be traced back to the international equivalent primary standard in the laboratory. The method is helpful for the cross verification of different types of field instruments at different sites and ensures the consistency of multiple field verification tests. It can also be used to calibrate satellite optical loads with high accuracy in the future.The field dissemination of the value of the quantity method can be divided into two partscalibration of a transfer spectroradiometer and field radiometric calibration. The influencing factors must be considered when a transfer spectroradiometer is used to calibrate the field spectroradiometer because of the differences between the laboratory and the actual site, such as relative spectral, radiation level, and temperature and humidity differences. Mathematical models are established to quantify the influence of various parameters on the spectral measurement accuracy, and a stray light correction model is built using the laser and filter method to correct the stray light caused by the relative spectral difference. The integrating sphere light source addition method is used to evaluate the nonlinearity at different radiation levels. In addition, whether spectral responsivity and detector temperature have a one-to-one correspondence is analyzed, and a temperature correction model is built according to the variation trend of spectral responsivity. After measuring and correcting the influence of the difference between the laboratory and actual site, the transfer spectroradiometer is utilized to transfer the radiometric quantity to the field spectroradiometer by using the symmetrical placement and exchange measurement method.

mots-clés

field dissemination of the value of quantity;stray light correction model;temperature correction model;exchange measurement;Uncertainty

References

  1. 1.
    Bouvet M, Thome K, Berthelot B, Bialek A, Czapla-Myers J, Fox N P, Goryl P, Henry P, Ma L L, Marcq S, Meygret A, Wenny B N and Woolliams E R. 2019. RadCalNet: a radiometric calibration network for earth observing imagers operating in the visible to shortwave infrared spectral range. Remote Sensing, 11(20): 2401
  2. 2.
    Czapla-Myers J S, Thome K J, Cocilovo B R, McCorkel J T and Buchanan J H. 2008. Temporal, spectral, and spatial study of the automated vicarious calibration test site at Railroad Valley, Nevada//Proceedings of SPIE 7081, Earth Observing Systems XIII. San Diego: SPIE: 70810I
  3. 3.
    General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China. 2016. JJF 1033-2016 Rule for the examination of measurement standards. Beijing: China Quality Inspection Press
  4. 4.
    Gou Z Y, Yan L, Chen W, Jing X, Yin Z Y and Duan Y N. 2012. In-flight absolute radiometric calibration of UAV hyperspectral camera and its validation analysis. Spectroscopy and Spectral Analysis, 32(2): 430-434
  5. 5.
    Kübarsepp T, Haapalinna A, Kärhä P and Ikonen E. 1998. Nonlinearity measurements of silicon photodetectors. Applied Optics, 37(13): 2716-2722 .
  6. 6.
    Li C R, Ma L L, Tang L L, Gao C X, Qian Y G, Wang N and Wang X H. 2021. A comprehensive calibration site for high resolution remote sensors dedicated to quantitative remote sensing and its applications. Journal of Remote Sensing, 25(1): 198-219
  7. 7.
    Li L, Dai C H, Wu Z F and Wang Y F. 2019. Laboratory calibration and temperature research of spectral radiometric instruments. Spectroscopy and Spectral Analysis, 39(6): 1965-1969
  8. 8.
    Li X, Zheng X B and Yin Y P. 2014. Progress in automated site vicarious calibration technologies. Journal of Atmospheric and Environmental Optics, 9(1): 17-21
  9. 9.
    Ma L L, Zhao Y G, Woolliams E R, Dai C H, Wang N, Liu Y K, Li L, Wang X H, Gao C X, Li C R and Tang L L. 2020. Uncertainty analysis for RadCalNet instrumented test sites using the Baotou sites BTCN and BSCN as examples. Remote Sensing, 12(11): 1696
  10. 10.
    Meftah M, Damé L, Bolsée D, Hauchecorne A, Pereira N, Sluse D, Cessateur G, Irbah A, Bureau J, Weber M, Bramstedt K, Hilbig T, Thiéblemont R, Marchand M, Lefèvre F, Sarkissian A and Bekki S. 2018. SOLAR-ISS: a new reference spectrum based on SOLAR/SOLSPEC observations. Astronomy and Astrophysics, 611: A1
  11. 11.
    Meygret A, Santer R P, and Berthelot B. 2011. ROSAS: a robotic station for atmosphere and surface characterization dedicated to on-orbit calibration//Proceedings of SPIE 8153, Earth Observing Systems XVI. San Diego: SPIE: 815311 [DOI: 10.1117/12.892759]
  12. 12.
    Price L L A, Hooke R J and Khazova M. 2014. Effects of ambient temperature on the performance of CCD array spectroradiometers and practical implications for field measurements. Journal of Radiological Protection, 34(3): 655-673
  13. 13.
    Salim S G R, Fox N P, Theocharous E, Sun T and Grattan K T V. 2011. Temperature and nonlinearity corrections for a photodiode array spectrometer used in the field. Applied Optics, 50(6): 866-875
  14. 14.
    Santer R P, Six C and Buis J P. 2003. Vicarious calibration on land site using automatic ground-based optical measurements: applications to SPOT-HRV//Proceedings of SPIE 4891, Optical Remote Sensing of the Atmosphere and Clouds III. Hangzhou: SPIE: 524-534
  15. 15.
    Shin D J, Park S, Jeong K L, Park S N and Lee D H. 2014. High-accuracy measurement of linearity of optical detectors based on flux addition of LEDs in an integrating sphere. Metrologia, 51(1): 25-32
  16. 16.
    Thompson A and Chen H M. 1994. Beamcon III, a linearity measurement instrument for optical detectors. Journal of Research of the National Institute of Standards and Technology, 99(6): 751-755.
  17. 17.
    Wan Z, Li B Y, Li X S, Liu Z X, Liu H X and Ye Z. 2012. Influence of integrating sphere source’s distribution temperature on broadband optical remote sensor’s absolute radiometric calibration and correcting methods. Chinese Journal of Luminescence, 33(8): 863-868
  18. 18.
    Wu Z F, Dai C H, Wang Y F and Zou J H. 2016. Measurement of optical parameters for array spectroradiometers. Optics and Precision Engineering, 24(8): 1902-1907
  19. 19.
    Xu M M, Jiang Q W, Liu W Q, Zhang Y C, Chen J X and Zeng Y. 2014. An improved method for optical system design and optimization of double grating spectrometer. Infrared and Laser Engineering, 43(1): 184-189
  20. 20.
    Zheng X B. 2011. High-accuracy radiometric calibration of satellite optical remote sensors. Spacecraft Recovery & Remote Sensing, 32(5): 36-43
  21. 21.
    Zong Y Q. 2010. Uncertainty analysis of stray-light correction//Proceedings of 2010 CIE Expert Symposium on Spectral and Imaging Methods for Photometry and Radiometry. [s.l.]: [s.n.]
  22. 22.
    Zong Y Q, Brown S W, Johnson B C, Lykke K R and Ohno Yoshi. 2006. Simple spectral stray light correction method for array spectroradiometers. Applied Optics, 45(6): 1111-1119

Lire l'article complet